W.L. GORE & ASSOCIATES GORE DUALMESH® PLUS BIOMATERIAL; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number 1DLMCP02
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).(b)(6).It should be noted that the gore® dualmesh® plus biomaterial instructions for use addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.The gore® dualmesh® plus biomaterial instructions for use also states: ¿strict aseptic techniques should be followed.If an infection develops, it should be treated aggressively.An unresolved infection may require removal of the material.
 
Event or Problem Description
It was reported to gore that the patient underwent incisional hernia repair on (b)(6) 2010 whereby a gore® dualmesh® plus biomaterial was implanted.The complaint alleges that on (b)(6) 2014 an additional procedure occurred whereby the gore device was explanted.It was reported the patient alleges the following injuries: mesh removal, mesh was detached and wadded up and chronically infected.Additional event specific information was not provided.
 
Additional Manufacturer Narrative
H6: updated results code.Conclusion code remains unchanged.
 
Additional Manufacturer Narrative
H6: code 4316 (appropriate term/code not available) is being used for: duplicate complaint.See mfr report #3003910212-2019-00445.
 
Additional Manufacturer Narrative
H6: updated investigation conclusions: d17: appropriate code/term not available for ¿false claim¿ and ¿withdrawn.¿ h6: health effect impact code: f26: no health consequences or impact.H6: medical device component: g07002: appropriate term not available for false claim.This claim was withdrawn, and the alleged product complaint is no longer being pursued at this time.No further information was received to substantiate the allegations of ¿mesh removal¿ and ¿additional surgery¿ thus the investigation findings do not lead to a clear conclusion about the cause of the reported adverse event and the device complaint or problem cannot be confirmed.Therefore, this event is being coded as no clinical signs, symptoms or conditions, no health consequences or impact and will be closed as false claim, withdrawn and no problem detected.It should be noted that the gore® dualmesh® plus biomaterial instructions for use include warnings and addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.¿ as with any surgical procedure, there are always risks of complications for surgical repair of hernias and soft tissue deficiencies, with or without mesh.These may include but are not limited to, adhesions and related harms, bleeding, bowel obstruction, compromised device biocompatibility, contamination which may lead to patient harms, device damage, dysphagia, erosion or extrusion and related harms, exposure or protrusion and related harms, fever, fistula, gerd recurrence, defect recurrence and related harms, ileus, increased procedure time and related harms, irritation or inflammation, infection, mesh migration, mesh contraction, pain, paresthesia, perforation, revision/re-intervention, seroma or hematoma and related harms, tissue ischemia, wound complications and wound dehiscence and additional intervention including surgery.Many of the potential complications are associated with the patient¿s underlying disease progression, co-morbidities, additional medical history and/or other surgical procedures.The above inherent risks are typically detailed in standard informed consent documents.The device was not able to be returned to gore for evaluation; therefore, a direct product analysis could not be conducted.The status of the device is unknown as no record of explant was provided.Review of the manufacturing and sterilization records verified that the lot met all pre-release specifications.Section c1: name: plus antimicrobial product coating.Manufacturer/compounder: w.L.Gore & associates, inc.Lot number: 7349388.Additional manufacturer narrative: the plus antimicrobial product coating contains silver carbonate [approximately 800 micrograms per cubic centimeter of product (g/cm3)], and chlorhexidine diacetate [approximately 1600 micrograms per cubic centimeter of product (g/cm3)].W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE DUALMESH® PLUS BIOMATERIAL
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
MPD APC B/P
p.o. box 1408
elkton MD 21922 1408
Manufacturer Contact
heidi womble
9285263030
MDR Report Key9379375
Report Number3003910212-2019-00489
Device Sequence Number16277384
Product Code FTL
UDI-Device Identifier00733132601097
UDI-Public00733132601097
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K063435
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2014
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2012
Device Model Number1DLMCP02
Device Catalogue Number1DLMCP02
Device Lot Number7349388
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Initial Report FDA Received Date11/26/2019
Supplement Report FDA Received Date12/16/2019
01/16/2020
06/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-