| Model Number |
ZXR00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided; best estimate is between (b)(6) 2019 and (b)(6) 2019.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxr00 24.5 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2019.It was later explanted on (b)(6) 2019 due to blurry vision.The replacement lens is a non-johnson and johnson lens and the patient was reported to be doing well post exchange.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.: returned to manufacturer on: 11/25/2019.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site in a little jar.The product was disinfected and visual inspection with the unaided eye shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be confirmed.The complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no similar complaints were received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|