• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARTICULAR SURFACE WITH HINGE POST EXTENSION SIZE E 12 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARTICULAR SURFACE WITH HINGE POST EXTENSION SIZE E 12 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Swelling (2091); Tissue Damage (2104); Limited Mobility Of The Implanted Joint (2671)
Event Date 09/30/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001822565-2019-04360.Concomitant medical products: femoral component size e left item# 00588001501 lot# 63384340, stem extension straight 15mm dia x 100mm length item# 00598801015 lot# 63464365.Report source: (b)(6).The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient was revised approximately two and half years post implantation due to implant fracture.Patient experienced left knee pain and difficulty ambulating.Left knee revision performed, noted the hinge had fractured and fragment had migrated, along with poly fracture.Patient had tendon tear repaired with anchors.There is no additional information at this time.There is no additional information at this time.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the returned product identified fracture hinge and bone cement is adhering to the femur.Articular surface is damaged in the anterior side.Fractured hinge was submitted for further analysis.Analysis determined that bending overload caused the fracture.The crack was suspected to be initiated near the inner diameter and propagated towards the outer diameter of the hinge.Fractured surface appear to be excessive smeared edges and biological debris.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: no complication was noted during both initial and revision surgery.The patient presented with patella tendon rupture along with pain, infra-articular serous effusion.Ct scan noted hinge fracture and metal fragments were migrated.Large amount of poly debris reaction along the anteromedial area was noted and poly was fractured along the anterior side.Scar tissues were noted.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.A definitive root cause cannot be determined.Per package insert 87-6203-904-23 rev c, nexgen rotating hinge knee: fracture, pain, swelling, wear is known adverse effect of this system.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE WITH HINGE POST EXTENSION SIZE E 12 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9380429
MDR Text Key168222223
Report Number0001822565-2019-05071
Device Sequence Number1
Product Code KRO
UDI-Device Identifier08809024199361
UDI-Public(01)08809024199361
Combination Product (y/n)N
PMA/PMN Number
K013385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberN/A
Device Catalogue Number00588005012
Device Lot Number63516115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/04/2019
Initial Date FDA Received11/26/2019
Supplement Dates Manufacturer Received03/30/2020
Supplement Dates FDA Received04/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight120
-
-