• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. WORKMATE CLARIS SYSTEM DISPLAY PLUS; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. WORKMATE CLARIS SYSTEM DISPLAY PLUS; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H700124
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
During a power outage, the cpu would not boot back into the program.Despite several power cycling attempts.The procedure was allegedly cancelled.
 
Event or Problem Description
During a power outage, the cpu would not boot back into the program, despite several power cycling attempts.The procedure was beyond the therapy stage so no additional recording or ablation was done after this event.The case was ended and the patient was discharged in stable condition.The patient needed to be rescheduled for a second part of the procedure which was not attempted after the system went down.The system was rebooted and eventually came back online.A ups was installed to resolve the issue.
 
Additional Manufacturer Narrative
Additional information: b5, g4, g7, h2, h3, h6.A review of the device history record was not possible since the system serial number was provided and the component serial number was not.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKMATE CLARIS SYSTEM DISPLAY PLUS
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key9382901
Report Number2184149-2019-00237
Device Sequence Number18952490
Product Code DQK
UDI-Device Identifier05415067001245
UDI-Public05415067001245
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH700124
Device Catalogue NumberH700124
Device Lot Number6044048
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/28/2019
Supplement Date Received by Manufacturer01/30/2020
Initial Report FDA Received Date11/27/2019
Supplement Report FDA Received Date02/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-