The country of origin is (b)(6).Occupation is lay user/patient.The complained test strips have been calibrated against the who standard rtf/16 and affected by recall.The corresponding retention test strips were tested in comparison to masterlot #28632180 (recalibrated lot to rtf/09).The corresponding retention material complies with the specification, based on requirements of the regular retention testing process of the qc department.The retention material and comparable masterlot test strips did not show abnormalities.Investigations have indicated results are reliable from 0.8 to 4.5 inr, since the calibration data covers this measuring range very well.Retention test strips were tested in comparison with the master lot coaguchek xs pt (ml 13 #286321-80 recal.Rtf/09).For this purpose, two human blood samples from marcumar donors and two internal reference meters were used.Retention samples were acceptable.No error messages occurred.
|