Catalog Number 150600009
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 06/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient underwent a second right knee revision due to pain, instability, and tibial tray loosening at the cement to implant interface.The patella component was not revised.Depuy cement was used during the primary operation.Doi: (b)(6) 2014 tibial insert, tibial tray, femoral component, and depuy cement; doi: (b)(6) 2018 patella component; dor: (b)(6) 2018.Right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 9 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9384393
Report Number1818910-2019-118423
Device Sequence Number10882527
Product Code JWH
UDI-Device Identifier10603295042433
UDI-Public10603295042433
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number150600009
Device Lot Number7779955
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/01/2019
Supplement Date Received by Manufacturer12/19/2019
Initial Report FDA Received Date11/27/2019
Supplement Report FDA Received Date12/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-