HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - OUTFLOW GRAFT; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1125
Medical Device Problem Code Misassembly by Users (3133)
Health Effect - Clinical Codes Pleural Effusion (2010); Blood Loss (2597)
Date of Event 11/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that three days post ventricular assist device (vad) implant, the patient was noted to have a left chest effusion on x-ray.The patient also had increased central venous pressure, worsening right ventricular function, and elevated pulmonary artery pressure.The patient was taken back to the operating room for exploration.Approximately 300 milliliters of blood was removed and it was found that the outflow graft (ofg) anastomosis site was leaking.Additional sutures were put in place to reinforce the anastomosis with the ascending aorta.The ofg remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the outflow graft was not returned for evaluation.Log file analysis could not be performed since log files c overing the reported event date were not available for analysis.The reported event could not be confirmed due to insufficient evidence.There is no evidence to suggest that a device malfunction caused or contributed to the reported event.Based on the available i nformation, a possible root cause of the reported "outflow graft anastomosis site leaking" event including but not limited to operator's technique during implant.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - OUTFLOW GRAFT
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key9385618
Report Number3007042319-2019-14367
Device Sequence Number14081973
Product Code DSQ
UDI-Device Identifier00888707000222
UDI-Public00888707000222
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2023
Device Model Number1125
Device Catalogue Number1125
Device Lot Number17456868-1392
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/21/2019
Supplement Date Received by Manufacturer01/15/2020
Initial Report FDA Received Date11/27/2019
Supplement Report FDA Received Date01/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD; 1103 VAD
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age52 YR
Patient Weight90
-
-