| Model Number |
PED-500-25 |
| Medical Device Problem Codes |
Activation Failure (3270); Migration (4003); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pipeline flex has not been returned for evaluation; product analysis cannot be performed.The device was not returned; the reported event could not be confirmed.The cause of the event could not be conclusively determined from the reported information.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that the pipeline fell into the aneurysm while attempting to retrieve the delivery system back into the microcatheter.Prior to the event, the microcatheter was used to deliver the pipeline to the aneurysm.The pipeline was then released, but could not be retrieved causing it to fall into the aneurysm.The microcatheter was removed and the distal part was observed to be stretched and accordioned.Another pipeline was implanted to bridge the connection.No patient injury was reported as a result of the event.The aneurysm diameters were 24.1mm/19.8mm.The aneurysm was not ruptured and saccular.
|
| |
|