• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-500-25
Medical Device Problem Codes Activation Failure (3270); Migration (4003); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pipeline flex has not been returned for evaluation; product analysis cannot be performed.The device was not returned; the reported event could not be confirmed.The cause of the event could not be conclusively determined from the reported information.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that the pipeline fell into the aneurysm while attempting to retrieve the delivery system back into the microcatheter.Prior to the event, the microcatheter was used to deliver the pipeline to the aneurysm.The pipeline was then released, but could not be retrieved causing it to fall into the aneurysm.The microcatheter was removed and the distal part was observed to be stretched and accordioned.Another pipeline was implanted to bridge the connection.No patient injury was reported as a result of the event.The aneurysm diameters were 24.1mm/19.8mm.The aneurysm was not ruptured and saccular.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key9387257
Report Number2029214-2019-01179
Device Sequence Number13959737
Product Code OUT
UDI-Device Identifier00847536019751
UDI-Public00847536019751
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/06/2022
Device Model NumberPED-500-25
Device Lot NumberA779086
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2019
Initial Report FDA Received Date11/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-