| Model Number |
110880 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
Extubate (2402)
|
| Date of Event |
11/10/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the unit ventilation volume decrease alarm was generated and cuff check was performed; however, the cuff pressure was appropriate.Customer stated tube abnormality was suspected, although placement position was checked with fiber, there was no problem, the cuff leak would not stop, and water was also being pooled in the inflation line heavily, it was judged that the condition was unmanageable and the tube was replaced.Reintubation was required.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: one sample of device was returned for evaluation.An inflation test was performed and it inflated without any issue.The returned unit was leak tested under water and no leakage was detected.The returned unit remained inflated for a 24- hour period and no leakage was detected.The reported defect could not be detected on the unit returned for evaluation.All of the products undergo a four hour inflate/deflate test and it is at this stage of the process that any defects are removed from the lot.The most probable root cause is an inflation device remained in the inflation valve allowing the air to escape.Instructions for use (ifu) states: syringes, three-way stopcocks or other cuff inflation devices should not be left inserted in the inflation valve for extended periods of time.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|