Model Number MMT-1714K
Device Problems Device Markings/Labelling Problem (2911); Power Problem (3010)
Patient Problem Shock, Insulin (2264)
Event Date 11/23/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
It was reported that their insulin pump had a open book image.The customer¿s blood glucose was unknown.Troubleshooting was performed for open book image.Insulin pump had no pump error before critical pump error.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device received with critical pump error due to corroded electronic assemblies.Unable to perform self test, rewind, seating, basic occlusion test, occlusion test, force sensor test or displacement test due to critical pump error.Device received with corroded battery tube, corroded motor home switch and faded serial number label.The p-cap does lock properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1714K 630G BLACK MMOL CANADA
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9389547
MDR Text Key168853569
Report Number2032227-2019-124322
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169782396
UDI-Public(01)00643169782396
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1714K
Device Catalogue NumberMMT-1714K
Device Lot NumberHG22UW2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received04/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-