| Model Number |
7209411 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during acl reconstruction surgery, the 8x35mm was opened but it was a 9x35mm in the packaging.The device was used on patient.A backup device was available to complete the procedure with no significant delay or patient injuries.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
One 7209411 rci 8x35mm screw used for treatment, was returned for evaluation.The diameter of the item returned for evaluation measured 9mm.Complaint history review indicated no similar allegations for the lot number reported.Dhr/batch/lot review did not indicate a condition or product/procedure failure that supported the allegation.Although we cannot determine when or where the product was put into an 8mm shelf carton, engineering has been made aware.At this time, this is an isolated incident.If relevant information becomes available to assist with further evaluation, the complaint may be revisited.No further actions required at this time.
|
| |
|