| Model Number |
2107-36 |
| Medical Device Problem Code |
Failure to Interrogate (1332)
|
| Health Effect - Clinical Codes |
Syncope (1610); Dyspnea (1816)
|
| Date of Event |
11/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 2938836-2019-16805.It was reported that the patient presented in the emergency department (ed) with near syncope and dyspnea.A merlin transmission was attempted without success.The device was unable to be interrogated.The patient stated being non-compliant for routine device follow-ups and noted the device vibrating approximately two years prior to this (b)(6) visit.The physician assumed this vibratory alert was for eri.During the generator change-out on a device suspected to be at eos, externalized conductors were observed on the riata lead via x-ray.The device was explanted and replaced.The rv lead was capped and replaced on (b)(6) 2019.The patient was stable throughout the procedure and remained stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported field event of interrogation failure was confirmed in the lab.A longevity calculation was performed using default parameters and showed that the battery depletion was normal.The device was tested on the bench and no anomalies were found.The cause of the communication failure was determined to be low battery voltage due to normal battery depletion.
|
| |
|