| Model Number |
0500308E |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
A user facility nurse reported that a dialyzer blood leak occurred 40 minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak.The machine, a fresenius 4008s machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 3 ml there was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be discarded and not available to be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.The nurse provided a photo of the dialyzer with blood on the outside of the fiber bundle inside the dialyzer.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.
|
| |
|
Search Alerts/Recalls
|