• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10604
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.
 
Event or Problem Description
It was reported that device contamination occurred.A 3.00x24mm synergy stent was used in a procedure.However, the stent was contaminated.No patient complications were reported and the patient's status was stable.
 
Event or Problem Description
It was reported that device contamination occurred.A 3.00x24mm synergy stent was used in a procedure.However, the stent was contaminated.No patient complications were reported and the patient's status was stable.
 
Additional Manufacturer Narrative
Device is a combination product.Device evaluated by mfr.:synergy ii us mr 3.00 x 24 mm stent delivery system was returned for analysis in a biohazard bag, without product mandrel & stent protector and with the proximal region of the hypotube shaft including the manifold missing.A visual examination of the stent found signs of stent damage.Stent struts in mid to distal stent region were noted to be lifted and pulled over the distal balloon cone.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of damage on the distal end of the bumper tip.A visual and tactile examination of the hypotube shaft found the shaft broken at 50.4 cm proximal to the distal end of the tip as well as a shaft kink situated at 47.2 cm proximal to the distal end of the tip.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.D10: returned to manufacturer date corrected from 22nov2019 to 05dec2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9395344
Report Number2134265-2019-14229
Device Sequence Number18927853
Product Code NIQ
UDI-Device Identifier08714729840237
UDI-Public08714729840237
Combination Product (Y/N)N
Initial Reporter CountryRQ
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10604
Device Catalogue Number10604
Device Lot Number0023943021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2019
Initial Date Received by Manufacturer 11/06/2019
Supplement Date Received by Manufacturer01/03/2020
Initial Report FDA Received Date12/02/2019
Supplement Report FDA Received Date01/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-