MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF XLPE 28 0 DEG 46-48 SZ D; PRSTHESIS, HIP, SEMI-CONSTRAINED, MTAL/POLYMER, PRUS UNCMNTD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF XLPE 28 0 DEG 46-48 SZ D; PRSTHESIS, HIP, SEMI-CONSTRAINED, MTAL/POLYMER, PRUS UNCMNTD Back to Search Results
Model Number 71333373
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that revision surgery was performed due to dislocation, because of wear.The patient requires abrasion loss of the liner.
 
Additional Manufacturer Narrative
It was reported that revision surgery was performed due to dislocation, because of wear.The affected devices, used in treatment, were returned and evaluated.A lab analysis conducted during this investigation noted that examination of the od of the shell revealed tissue on growth which appears to be bone.The inside of the shell does not exhibit any unexpected wear or scratches.The femoral head does not display any markings of damage resulting from a dislocation.The hole cover and screws do not have any noted physical changes which would indicate they played a part in a dislocation.Examination of the od of the poly does not show any abnormal wear.A yellow tint can be seen, this likely is due to absorption of bio-logical fluids over the time of implantation.The id does not show any abnormal wear or marks of neck impingement that would be expected in the presence of dislocation.The head appeared to be tight with no evidence of wear as reported.This visual analysis of the returned devices does not support the reported statement of dislocation due to wear.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of complaint history on the listed parts revealed no prior complaints for the listed batch with the same failure mode.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A relationship between the devices and the reported incident could not be corroborated.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Per e-mail communication the devices had been implanted for 10 years.Possible causes could include but not limited to lifetime of device or device friction.Based on this investigation, the need for corrective action is not indicated.We will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF XLPE 28 0 DEG 46-48 SZ D
Common Device Name
PRSTHESIS, HIP, SEMI-CONSTRAINED, MTAL/POLYMER, PRUS UNCMNTD
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9397079
Report Number1020279-2019-04234
Device Sequence Number3095960
Product Code LPH
UDI-Device Identifier03596010455857
UDI-Public03596010455857
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K002747
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date10/27/2016
Device Model Number71333373
Device Catalogue Number71333373
Device Lot Number06KM02575
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2019
Initial Date Received by Manufacturer 11/07/2019
Supplement Date Received by Manufacturer02/18/2020
Initial Report FDA Received Date12/02/2019
Supplement Report FDA Received Date02/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PART NO.: 11915 / LOT NO.: 07DM19709; PART NO.: 71330001 / LOT NO.: UNKNOWN; PART NO.: 71336048 / LOT NO.: 07DM11210; PART NO.: 71336525 / LOT NO.: UNKNOWN (2 DEVICES)
Outcome Attributed to Adverse Event Required Intervention;
-
-