| Model Number |
1504-11-206 |
| Medical Device Problem Codes |
Noise, Audible (3273); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Adhesion(s) (1695); Edema (1820); Pain (1994); No Code Available (3191)
|
| Date of Event |
10/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > the device associated with the reported event was not returned to the investigation site for evaluation.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
(b)(6).Clinical adverse event received for hypersensitive scar knee.Event is not serious and is considered mild.Event is not related to device and is definitely related to procedure.Date of implantation: (b)(6) 2019, date of event (onset): (b)(6) 2019, (right knee).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added h6 (patient code).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H6 patient code: no code available (3191) used to capture the patient code minor injury.
|
| |
|
Search Alerts/Recalls
|