COVIDIEN MFG SOLUTIONS S.A. ULTRAFLO; CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number 8888221220
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intraoperatively, the clamp was opened when the catheter was installed on the venous side of the catheter.Catheter was not repaired, there was no leak, no luer adapter issue, no patient symptoms or complications with event, medical or surgical intervention was not needed, did not lead to patient hospitalization.There was no reported patient injury.
 
Additional Manufacturer Narrative
Additional info: b5, g4, h3, h6.H3 evaluation summary: post market vigilance (pmv) led a photographic evaluation of one device.A visual inspection of the returned photo noted a catheter product with the venous clamp unclamped.No abnormalities were noted.As no physical product was returned, functional testing could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intraoperatively, the clamp was opened when the catheter was installed on the venous side of the catheter.Catheter was not repaired, tego was not utilized, there was no leak, no luer adapter issue, no patient symptoms or complications with event, medical or surgical intervention was not needed, did not lead to patient hospitalization, nothing unusual observed prior to use, blood transfusion done, clorhexidina was used on the insertion site prior to product placement.Catheter was changed to complete treatment.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLO
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key9397789
Report Number3009211636-2019-00256
Device Sequence Number2763989
Product Code MPB
UDI-Device Identifier10884521125735
UDI-Public10884521125735
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K120674
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2023
Device Model Number8888221220
Device Catalogue Number8888221220
Device Lot Number8888221220
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/12/2019
Supplement Date Received by Manufacturer11/28/2019
Initial Report FDA Received Date12/02/2019
Supplement Report FDA Received Date01/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-