| Model Number |
N560 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the device does not mark the digits well.There was no allegation of patient death or serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g4, h3, h6 h3 evaluation summary: one device was received for evaluation.The reported event was confirmed.The device failed to meet specification as it was received or made available for evaluation.The front housing and top cover are damaged.The equipment is ready to use and safe for the patient according to the manual manufacturer's standards.Functional test with approved success.Battery must be replaced every two years as indicated in the service manual.The investigation of observed condition isolated the failure to the damaged housing and old battery.No new formal investigation is required, the event will be included in trending and monitoring.If information is provided in the future, a supplemental report will be issued.
|
| |
|