| Model Number |
0502990070 |
| Medical Device Problem Code |
Optical Obstruction (3002)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there could be a potential of adverse consequences due to image loss.
|
| |
|
Event or Problem Description
|
|
It was reported that there could be a potential of adverse consequences due to image loss.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: can not see through scope.Confirmed failure: fiber/cone damage at sidearm,outer tube damaged (bent, dented),broken rod lens,loose prism.Probable root cause: ¿ improper cleaning process at assembly; ¿ loss of in-joint pressure (arthroscopes only); ¿ loss of insufflation (laparoscopes only); ¿ shipping damage; ¿ use error.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|