MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE MODERATE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3542911 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/26/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old patient underwent a breast augmentation primary with a mentor siltex contour profile moderate 225cc and experienced left implant deflation.As a result, the patient had undergone removal and replacement with saline mentor breast implant of unknown type on (b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
On 12/10/19, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 1/28/2020, during visual inspection of the device it was observed with no apparent damage.Leak testing revealed there was leakage from the valve.Excessive manipulation may have caused the valve leak.The ifu states ¿not to manipulate (i.E., squeeze) the valve excessively, which may cause valve leakage¿.However, this cannot be conclusively determined.The manufacturing records were reviewed, and the manufacturing criteria were met prior to the release of this lot.The second product received is related to a concomitant device, there are neither deficiencies alleged nor patient injuries.Therefore no further investigation is required.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information will be included in complaint trending that is reviewed and monitor by quality assurance on a regular basis to determine if further action is necessary.Manufacturer¿s reference number: (b)(4).
|
| |
|