MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL GALAXY G3 XSFT HEL 2MM X 4CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL GALAXY G3 XSFT HEL 2MM X 4CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number GLX120204
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.The product is available for evaluation and testing; however, it has not been received to date.Additional information will be submitted within 30 days of receipt.This is one of three products involved with the complaint and the associated manufacturer report numbers are 3008114965-2019-01260 and 3008114965-2019-01262.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Event or Problem Description
As reported by a healthcare professional, during a coil embolization of a ruptured posterior communicating artery (pcom) aneurysm of roundabout 2mm, the first coil was too big, then a 2mm x 4cm (glm920040, l11688) galaxy g3 mini was chosen.It was perfectly configurated in the aneurysm, so they wanted to detach, however, the system did not detach when using a enpower dcb 2 (dcb2000500, l12008) and an enpower control cable (ecb00018200, l11342).They used another enpower dcb 2 (dcb2000500, l12008) and another enpower control cable (ecb00018200, l11342, but it did not detach.A 1mm x 1cm galaxy g3 mini (glm910010, l11221) was attempted to be detached but failed.And a 1.5mm x 2cm galaxy g3 mini (glm915020, l12474) also failed to detach.In total they tried 3 different coils, all of them did not detach or did not show the connection signal.They changed the cables and boxes, but nothing worked.In total two dcb and two cables were used.The procedure was completed with competitor¿s coils.There was no report of the event resulting in prolongation of the procedure.Adequate flush had been maintained.The devices were prepped as per the instructions for use (ifu).Pre-deployment electrical testing was performed with the first coil, when it was deployed.The other coils had been already tested when they unpacked them.One of the dcb¿s was used for a former coiling and everything worked.The coils were still attached to the coil delivery system when removed from the patient.There is no information as to the coil being stretched or damaged when removed.It was not necessary to remove concomitant devices with the complaint devices.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by a healthcare professional, during a coil embolization of a ruptured posterior communicating artery (pcom) aneurysm of roundabout 2mm, the first coil was too big, then a 2mm x 4cm (glm920040, l11688) galaxy g3 mini was chosen.It was perfectly configurated in the aneurysm, so they wanted to detach, however, the system did not detach when using a enpower dcb 2 (dcb2000500, l12008) and an enpower control cable (ecb00018200, l11342).They used another enpower dcb 2 (dcb2000500, l12008) and another enpower control cable (ecb00018200, l11342, but it did not detach.A 1mm x 1cm galaxy g3 mini (glm910010, l11221) was attempted to be detached but failed.And a 1.5mm x 2cm galaxy g3 mini (glm915020, l12474) also failed to detach.In total they tried 3 different coils, all of them did not detach or did not show the connection signal.They changed the cables and boxes, but nothing worked.In total two dcb and two cables were used.The procedure was completed with competitor¿s coils.There was no report of the event resulting in prolongation of the procedure.Adequate flush had been maintained.The devices were prepped as per the instructions for use (ifu).Pre-deployment electrical testing was performed with the first coil, when it was deployed.The other coils had been already tested when they unpacked them.One of the dcb¿s was used for a former coiling and everything worked.The coils were still attached to the coil delivery system when removed from the patient.There is no information as to the coil being stretched or damaged when removed.It was not necessary to remove concomitant devices with the complaint devices.One non-sterile galaxy g3 xsft hel 2mm x 4cm unit was received inside of a pouch.The received device was visually inspected, and no damages were noted on it.Then a microscopic inspection was performed, and the distal outer sheath had not been softened indicating that the resistance heating coil had not been heated and the detachment process was not initiated.No other damages were observed.Also, the marker band was found at 39cm from distal end and it was found within specification.The resistance of the device was measured with multimeter and a lab sample enpower control cable.Resistance initially measured approximately 54.1o, which is within the specification range.The received unit galaxy g3 xsft hel 2mm x 4cm was connected to a lab sample enpower control cable and then connected to the detachment control box received.The power was turned on and the system ready light illuminated.A detachment cycle was initiated when the detach button was pressed.After that the embolic coil was detached from the device.The complaint reported by the customer ¿coil - failure to detach¿ could not be confirmed because the device was found within the specifications in the functional test.After that the embolic coil was detached from the device.Neither the analysis nor the mre suggest that the failure reported could be related to the manufacturing process.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided the evidence presented by the returned device; however, it is possible that clinical and procedural factors, including device manipulation and device interaction, may have contributed to the reported failure and damages on the returned system.Based on the event description, there was an attempt to detach the device with an enpower control cable and an enpower dcb 2 control box.However, based on the analysis of the device received, the distal outer sheath had not been softened indicating that the resistance heating coil had not been heated and the detachment process was not initiated.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY G3 XSFT HEL 2MM X 4CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
MDR Report Key9401438
Report Number3008114965-2019-01261
Device Sequence Number11414836
Product Code KRD
UDI-Device Identifier10886704077749
UDI-Public10886704077749
Combination Product (Y/N)N
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberGLX120204
Device Catalogue NumberGLX120204
Device Lot NumberL11688
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2020
Initial Date Received by Manufacturer 11/11/2019
Supplement Date Received by Manufacturer02/10/2020
03/03/2020
Initial Report FDA Received Date12/03/2019
Supplement Report FDA Received Date02/24/2020
03/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENPOWER CONTROL CABLE; ENPOWER DCB 2; GALAXY G3 MINI 1.5MM X 2CM; GALAXY G3 MINI 1MM X 1CM
-
-