• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG., AHTO IRRIGATION PUMP; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG., AHTO IRRIGATION PUMP; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED Back to Search Results
Model Number 0250070601
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code Burn, Thermal (2530)
Date of Event 11/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was a thermal event.
 
Additional Manufacturer Narrative
Alleged failure: the tower generated a warning signal on the isulation monitor, which was not recognized by the responsible sister, because she was not instructed in these signals according to her own statement.Shortly after the signal, a jet of flame occurred at the ahto (approx.10-30 seconds), causing the power cable and pump to fuse.The incident did not influence the surgical result, as the incident took place before the beginning of the surgery (the patient had just been pushed into the room).Surgical delay of 15min.Not longer.There have been no changes in the procedure of the surgery.Nobody was injured or impaired.Please see the following compoments which belong to the tower affected in this event: ref 250-070-601 sn: (b)(4).Computer sdc-hd-ref 240-050-888 sn: (b)(4).Transmitter synk wireless system sn: (b)(4).Light source x8000 sn: (b)(4).Printer sdp1000 sn: (b)(4).Monitor, tft visionelect hd sn: (b)(4).Insufflator 40 liter sn: (b)(4).Camera processor 1188hd sn: (b)(4).Camera head 1188210105 sn: (b)(4).The failure(s) identified in the investigation is consistent with the complaint record.Probable root causes could be the pump's ac inlet got some fluid ingress (ex.Saline, cleaning chemical) causing short circuit on the ac inlet.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was a thermal event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., AHTO IRRIGATION PUMP
Common Device Name
APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key9403724
Report Number0002936485-2019-00533
Device Sequence Number10909060
Product Code GCX
UDI-Device Identifier07613327052138
UDI-Public07613327052138
Combination Product (Y/N)N
PMA/510(K) Number
K042454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number0250070601
Device Catalogue Number0250070601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2020
Initial Date Received by Manufacturer 11/21/2019
Supplement Date Received by Manufacturer11/21/2019
Initial Report FDA Received Date12/03/2019
Supplement Report FDA Received Date03/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-