Model Number TV-AB2980-J
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Integrity Problem (2978)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Failure of Implant (1924)
Date of Event 11/04/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot history records related to the subject device referenced in this complaint record were reviewed for potential contributing factors for the failure mode(s) described in this event.A review of the device quality records showed that the device(s) demonstrated compliance to established procedures and specifications at the time of manufacture.
 
Event or Problem Description
An ovation ix abdominal stent graft system was implanted to treat an abdominal aortic aneurysm.Post polymer fill of the aortic body stent graft, the patient experienced hypotension and was successfully treated for an anaphylactic reaction per the device ifu.The case continued without additional incident.At the conclusion of the procedure the aneurysm was excluded.The patient will continue to be monitored.
 
Additional Manufacturer Narrative
A clinical assessment of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging.Several attempts were made to obtain medical records and medical images, but no response was received.As such, procedure related harms, event determination, off label conditions, related patient harms and patient disposition could not be assessed with medical records or images.The reported intraoperative hypotension and successfully treated for an anaphylactic reaction per the device ifu is most likely device related due to a suspected polymer leak.As identified in fs-0012, the root cause for most polymer leaks is a material weakness adjacent to the polymer fill channel which may become compromised during pressurization with liquid polymer.Since this device remains implanted, endologix is unable to conduct product evaluation to conclusively ascertain root cause.However, based on the investigation associated with this fsn, this is the most likely cause associated with this event.As identified in fs-0012, the root cause for most polymer leaks is a material weakness adjacent to the polymer fill channel which may become compromised during pressurization with liquid polymer.Since this device remains implanted, endologix is unable to conduct product evaluation to conclusively ascertain root cause.However, based on the investigation associated with this fsn, this is the most likely cause associated with this event.On 06aug2018, endologix issued a field safety notice (fs-0009) regarding the risk of polymer leak events with the ovation ix aortic body stent graft.On 06may2020, endologix issued a follow-up safety update (fs-0012) reaffirming treatment recommendations for patients who experience a polymer leak during implantation and providing updated information on the current rate of polymer leaks, the rate of clinical harms and root cause.The manufacturing lot evaluation confirmed all devices met specifications prior to release.No additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.Corrections: g1,2: contact office- contact has been updated h6: result code: remove code 3233 h6: conclusion code: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
AORTIC MAIN BODY
Manufacturer (Section D)
ENDOLOGIX (TRIVASCULAR)
3910 brickway blvd
santa rosa CA 95403
MDR Report Key9404697
Report Number3008011247-2019-00173
Device Sequence Number19385408
Product Code MIH
UDI-Device IdentifierM701TVAB2980J1
UDI-Public+M701TVAB2980J1/$$3221002FS093019085
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/02/2022
Device Model NumberTV-AB2980-J
Device Catalogue NumberTV-AB2980-J
Device Lot NumberFS093019-08
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/06/2019
Supplement Date Received by Manufacturer11/06/2019
Initial Report FDA Received Date12/03/2019
Supplement Report FDA Received Date08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FSN-0009, FSN-0012
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OVATION CUSTOM SEAL FILL POLYMER.; OVATION IX ILIAC LIMB (2).; OVATION CUSTOM SEAL FILL POLYMER; OVATION IX ILIAC LIMB (2)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-