| Model Number |
DTBB1D4 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 6935m62 lead; implanted: (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a lead integrity alert (lia) on the right ventricular (rv) lead for high rate non-sustained episodes and elevated short ventricular intervals.There was also a noise warning, oversensing, and low pacing impedance on the rv lead, which was noted to be fractured.The patient was inappropriately shocked due to the rv lead issue.The therapy portion of the rv lead was programmed off.It was noted that this lead to premature battery depletion on the device.The device and lead were explanted and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
| |
|