| Model Number |
10464525 |
| Medical Device Problem Codes |
Unexpected Therapeutic Results (1631); High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mdrs 2517506-2019-00452 and 2517506-2019-00453 were filed for the same event.The customer contacted the siemens customer care center (ccc) and reported elevated cardiac troponin i (tni) patient results were obtained on a dimension exl 200 instrument.Siemens is investigating the event.
|
| |
|
Event or Problem Description
|
|
Elevated a cardiac troponin i (tni) results were obtained on patient serum samples on a dimension exl 200 instrument.The results were reported to the physician(s) who questioned the result.The results obtained were considered correct based on the patient's previous history of a positive tni result.There are no known reports of patient intervention or adverse health consequences due to the elevated tni results.
|
| |
|
Additional Manufacturer Narrative
|
|
Mdr 2517506-2019-00454 was filed on 03-dec-2019.Mdrs 2517506-2019-00452 and 2517506-2019-00453 were filed for the same event.Additional information (04-feb-2020): siemens healthcare diagnostics headquarters support center (hsc) concluded their investigation of the discordant falsely elevated cardiac troponin i (tni) result.Hsc evaluated the information provided.Additional information was requested from the customer, including message logs, clot check data, chem data, filter data and loci data.The requested data was not provided.No product problem was identified.The instrument is operational.The cause of this event is unknown.The device is performing within specifications.No further evaluation is required.Section h6 has been updated to reflect the hsc investigation.Supplemental mdrs 2517506-2019-00452 and 2517506-2019-00453 were filed for the same event.
|
| |
|
Search Alerts/Recalls
|