It was reported that during a shoulder surgery, the pump was not working correctly.It was unable to maintain a flow rate or pressure.The device stopped to work throughout the case for prolonged periods, having a difficultly visualization and need for use of flush function several times.An extra 60-90 minutes were added to the case resulting in extra time of the patient under anesthesia.It resulted in greater than normal swelling in the shoulder as well.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H3,h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of fluctuating pressure could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 2 hour burn-in on wet station utilizing low and high pressure and flow settings.Raw and zero transducer readings were normal and well within specs during all functional tests.At no time during functional testing did pump over pressurize, fluctuate, or become intermittent.All functions perform as expected.Pressure and flow readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
|