The device used for treatment was returned for evaluation.
The reported problem was visually confirmed.
The snap-lock nut is sheared off at the collet.
In addition, a visual inspection was performed on the part beyond the reported complaint.
The snap locknut is the old version without the pin.
Although the reported problem was visually confirmed, a functional evaluation was performed and failed the handpiece test with a t1 value of 98.
The device met all specifications upon release into distribution.
A complaint history review for similar reported/confirmed complaints associated with the handpiece and failure mode(s) "error message(s)" identified similar events.
The most likely cause of this event is the mechanical failure of the snap lock.
As a part of corrective actions, more robust design of the snap lock has been released.
|