| Model Number |
MMT-723NAL |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that insulin pump alarmed motor error.The customer¿s blood glucose level was unknown.Customer reported that the drive support cap appeared normal.The customer was assisted with troubleshooting.The customer was advised that the pump will need to be replaced and was advised to discontinue use of the pump and revert to a back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test, rewind test, basic occlusion test, prime/a33 test, excessive no delivery test and occlusion test, no motor error alarm during testing noted.The motor was tested outside the device and passed.(b)(4).
|
| |
|