Model Number 2426-0007 |
Device Problems
Stretched (1601); Material Deformation (2976); Insufficient Information (3190)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.Patient demographics requested however not provided.Requested lot however not provided.
|
|
Event Description
|
The product was received as an unexpected return with a bulge in the silicone segment.Although requested there was no other additional event details or patient impact provided by the customer.
|
|
Manufacturer Narrative
|
The customer¿s report was confirmed during the visual inspection as a balloon at the top of the silicone pump segment.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection observed that the set's silicone segment area near the upper fitment was found to be softer/weaker as compared to the rest of the silicone segment.Further visual inspection of the silicone tubing confirmed that the tubing's concentricity was within specification(s).No other anomalies or evidence of damage were observed upon initial visual inspection.Functional testing could not be confirmed.Previously investigated complaints for this same failure mode determined that the ballooning is caused by excess pressure within the silicone tubing segment.The root cause for the source of the excessive pressure is unknown.The root cause of the customer's report could not be identified.
|
|
Event Description
|
The product was received as an unexpected return with a "bulge" in the silicone segment.Although requested, there was no other additional event details or patient impact provided by the customer.
|
|
Search Alerts/Recalls
|