(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Device received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm during testing noted.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime compromised force sensor system and excessive no delivery test or verify motor error alarm, and excessive no delivery alarm due to buttons not responding and complete broken reservoir tube lip.Device received with cracked case and broken battery tube threads.Motor passed motor test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|