• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 18B-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 18B-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Model Number 52-1483
Medical Device Problem Codes Fracture (1260); Naturally Worn (2988)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Bone Fracture(s) (1870); Joint Disorder (2373); No Code Available (3191)
Date of Event 07/26/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pfs and medical records received.After review of the medical records the patient was revised on (b)(6) 2016.The pfs indicated the patient was revised for dislocation and a fractured stem.The stem has already been reported on (b)(4).Part/ lot hasn't been provided at this time.Ppf has no new allegations reported.After review of medical records, the patient was revised to perform an excisional irrigation and debridement as well as an extended trochanteric osteotomy of the hip to address dislocation, extensive metallosis and damaged trunnion of the femoral component.Updated the implant date of the head and the details of the impacted products.Doi: (b)(6) 2015 (head), doi: (b)(6) 2016 (liner), dor: (b)(6) 2016 (left hip).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: investigative update: 29 / oct / 2021.X-ray images have been provided for review.No device associated with this report was received for examination.X-ray images have been provided for review.None of the allegations reported regarding the femoral sleeve component can be verified.Nothing indicative of a product problem is identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key9409781
Report Number1818910-2019-120095
Device Sequence Number19830163
Product Code LPH
UDI-Device Identifier10603295170631
UDI-Public10603295170631
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K934412
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number52-1483
Device Catalogue Number521483
Device Lot Number2931505
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/07/2019
Supplement Date Received by Manufacturer12/06/2021
Initial Report FDA Received Date12/04/2019
Supplement Report FDA Received Date12/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/27/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age21 YR
Patient SexFemale
Patient Weight70 KG
-
-