• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFCSIGMA INS TRL STABL 8MM SZ4; KNEE INSTRUMENT : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFCSIGMA INS TRL STABL 8MM SZ4; KNEE INSTRUMENT : TIBIAL TRIALS Back to Search Results
Catalog Number 961450
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the 2 tibial insert trials was broken during case.No surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : the device associated with this report was not received for examination.The investigation could not verify or identify any product contribution to the current reported event without the device to examine.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFCSIGMA INS TRL STABL 8MM SZ4
Common Device Name
KNEE INSTRUMENT : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9410413
Report Number1818910-2019-120162
Device Sequence Number3987570
Product Code LXH
UDI-Device Identifier10603295238652
UDI-Public10603295238652
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number961450
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/11/2019
Supplement Date Received by Manufacturer01/06/2020
Initial Report FDA Received Date12/04/2019
Supplement Report FDA Received Date01/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-