| Model Number |
MT22719-BLU |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that signal loss over one hour occurred.The product was evaluated.An external visual inspection was performed and passed.A charge and boot test was performed and passed.A receiver functional test was performed and passed.A receiver pairing test was performed and passed.A review of the share logs was performed and signal loss was not found within the investigation window.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|