• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT CEMENTED STEM SZ3 STD; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT CEMENTED STEM SZ3 STD; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 157003090
Medical Device Problem Codes Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Codes Unspecified Infection (1930); Necrosis (1971); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 03/31/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pfs, ppf, implant record and medical records received.Pfs has no allegation ppf alleges infection.After review of medical records, patient was revised to address infected right hip arthroplasty.Operative notes reported of some superficial fluid and necrotic debris.The cup came out without any difficulty and the stem was grossly loose.Doi: (b)(6) 2015 (head & liner) doi: (b)(6) 2004 (stem, cup, screws, centralizer, cement, eliminator) dor: (b)(6) 2016 right hip 7th revision please see pc-(b)(4) right hip 1st revision.Pc-(b)(4) right hip 2nd revision.Pc-(b)(4) right hip 3rd revision.Pc-(b)(4) right hip 4th revision.Pc-(b)(4) right hip 5th revision.Pc-(b)(4) right hip 6th revision.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT CEMENTED STEM SZ3 STD
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9410801
Report Number1818910-2019-120190
Device Sequence Number11496577
Product Code JDI
UDI-Device Identifier10603295059530
UDI-Public10603295059530
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K013352
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Device Explanted Year2016
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/15/2009
Device Catalogue Number157003090
Device Lot NumberYX5DL1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/08/2019
Supplement Date Received by Manufacturer12/20/2019
Initial Report FDA Received Date12/04/2019
Supplement Report FDA Received Date12/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient Weight61
-
-