| Catalog Number |
471987 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Blood Loss (2597); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: it was reported that the set detached and patient had 2ml of blood loss.
|
| |
|
Additional Manufacturer Narrative
|
|
Two (2) used samples were returned for evaluation.After a visual inspection was performed, it was noted that the two samples appeared to be connected to non-b.Braun valves.The two sets were tested per specification with passing results.The reported defect was unable to be confirmed through sample evaluation.If a sample and/or any additional pertinent information becomes available, a follow-up will be submitted.
|
| |
|