| Catalog Number |
Y464G |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: confirm the total qty involved with reported issue during this procedure? 3 devices are reported.Confirm the lot # for all the devices involved with reported issue during this procedure? reported lot# is pgm779 only.Were the remaining 3 noticed with reported issue before use on patient while dispensing/handling ? or during use on patient while suturing ? no further information is available.No further information will be provided.Note: events reported in 2210968-2019-90543, 2210968-2019-90544, and 2210968-2019-90545.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown surgery on an unknown date and suture was used.During the procedure, when trying to use the needle-suture, the suture was detached from the needle.There were no adverse consequences to the patient.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A manufacturing record evaluation was performed for the finished device lot pgm779, and no non-conformances were identified.
|
| |
|