| Catalog Number |
BAR201101 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been returned for analysis.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.No: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial fibrillation (afib) procedure with an ep diagnostic catheter dynamic xt steerable decapolar large and the deflection became stuck.During the procedure, the curve of the catheter was stuck in a deflected position with full tension on the curved tip and could not be relaxed.The device was changed to another catheter.The procedure was completed and was not altered as a result of this event.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a patient underwent an atrial fibrillation (afib) procedure with an ep diagnostic catheter dynamic xt steerable decapolar large and the deflection became stuck.The device was examined and subjected to full function testing and evaluating the deflection and curve capabilities of the device.The catheter was observed to deflect and return to its neutral position.The distal tip moves and is not sticking or stuck, it is steerable.There are indications the product was removed from its packaging and handled.Though the catheter was observed with a bend in its shaft, it passed all functional tests as it met the expected deflection and curvature for this model.The reported issue is not confirmed.A manufacturing record evaluation was conducted for lot# 2104429 and no non-conformances were identified.Manufacturer's ref.No: (b)(4).
|
| |
|
Search Alerts/Recalls
|