• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO SILVERGLIDE BIPOLAR STANDARD FORCEPS CUSHING, 21CM / 2.0MM TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO SILVERGLIDE BIPOLAR STANDARD FORCEPS CUSHING, 21CM / 2.0MM TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 6710210020
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/12/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during testing by the manufacturer sales representative at the user facility, pieces of insulation were found to be missing from the device, posing the risk of a material being lost in a surgical site.There were no adverse consequences related to this event.
 
Additional Manufacturer Narrative
Follow-up report submitted to document device evaluation results.
 
Event or Problem Description
It was reported that during testing by the manufacturer sales representative at the user facility, pieces of insulation were found to be missing from the device, posing the risk of a material being lost in a surgical site.There were no adverse consequences related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILVERGLIDE BIPOLAR STANDARD FORCEPS CUSHING, 21CM / 2.0MM TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key9413260
Report Number0001811755-2019-03944
Device Sequence Number19866053
Product Code GEI
UDI-Device Identifier04546540714909
UDI-Public04546540714909
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K992931
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number6710210020
Device Catalogue Number6710210020
Device Lot Number222285
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2019
Initial Date Received by Manufacturer 11/12/2019
Supplement Date Received by Manufacturer03/10/2020
Initial Report FDA Received Date12/04/2019
Supplement Report FDA Received Date03/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-