| Model Number |
ACE45E |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device batch number and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, the white tissue pad had fallen off.The fallen piece was retrieved by forceps and was disposed of.Changed to another one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #p93m6v.Investigation summary: the device was returned with the tissue pad damaged, melted, not all present and a portion was not returned.In addition, the tissue pad was attached to the clamp arm and not detached as reported by the customer.The device was connected to a test hand piece and a gen11 and the device did activate during functional testing.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Probable causes of tissue pad damage are applying pressure between the instrument blade and tissue pad without having tissue between them.Avoid activating the blade without tissue between the blade and tissue pad to prevent damage to the tissue pad.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.
|
| |
|