• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII CR DEEP FLEX ISRT S5-6 9M; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII CR DEEP FLEX ISRT S5-6 9M; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71421542
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that when the patient fell, he reopened the surgical scar, which was treated with a covering and became infected.He underwent further surgery to debride infected tissue, inspected and wash out to the joint and replaced the poly insert.The tibial and femoral components were examined and found to be intact and well adhered.No operative reports are available.
 
Additional Manufacturer Narrative
It was reported that the patient reopened the surgical scar when he fell, which became infected.The affected genesis ii cr deep flexion insert, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident or adverse event could not be corroborated.No medical documents were received for investigation.No further medical assessment can be performed at this time.The fact that the surgical scar was reopened after the fall could be determined as the root cause of the infection.Review of the instructions for use and the risk management files identified the reported failure as potential adverse events.Without the return of the actual product involved and no patient medical records available, our investigation of this report is inconclusive.Based on this investigation, the need for corrective action is not indicated.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII CR DEEP FLEX ISRT S5-6 9M
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9414240
Report Number1020279-2019-04288
Device Sequence Number19830338
Product Code JWH
UDI-Device Identifier03596010512475
UDI-Public03596010512475
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K041825
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71421542
Device Catalogue Number71421542
Device Lot Number18BM00110
Initial Date Received by Manufacturer 11/08/2019
Supplement Date Received by Manufacturer11/08/2019
Initial Report FDA Received Date12/04/2019
Supplement Report FDA Received Date03/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-