MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAH
Medical Device Problem Codes Mechanical Problem (1384); Obstruction of Flow (2423)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The initial report was submitted with the wrong aware date.The correct date is 25-nov-2019.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Additional Manufacturer Narrative
(b)(4).A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.No motor error alarm or unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Event or Problem Description
The customer reported via phone call that the insulin pump received motor error.The customer¿s blood glucose level was unknown.The customer reported that they received a motor error and stopped delivering insulin.The insulin pump will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9414405
Report Number2032227-2019-127461
Device Sequence Number1712077
Product Code OYC
UDI-Device Identifier00643169077706
UDI-Public(01)00643169077706
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAH
Device Catalogue NumberMMT-723NAH
Device Lot NumberA7723NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2019
Initial Date Received by Manufacturer 01/04/2019
Supplement Date Received by Manufacturer12/10/2019
Initial Report FDA Received Date12/04/2019
Supplement Report FDA Received Date12/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age34 YR
Patient Weight72
-
-