| Model Number |
MMT-723NAH |
| Medical Device Problem Codes |
Mechanical Problem (1384); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The initial report was submitted with the wrong aware date.The correct date is 25-nov-2019.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.No motor error alarm or unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump received motor error.The customer¿s blood glucose level was unknown.The customer reported that they received a motor error and stopped delivering insulin.The insulin pump will not be returned for analysis.
|
| |
|