(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Device received with intermittent button response due to flattened (escape, down and act ) button dome switch (no crease) and partial unlocked j2/lcd keypad connector noted during visual inspect.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime/compromised force sensor system and excessive no delivery test due to buttons not responding.Device received with cracked case display window corner.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|