(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Device received with intermittent button response due to flattened express bolus and act button dome switch (creased).No button error alarm during testing.No unlocked j2/lcd keypad connector.Device passed the displacement test, rewind, basic occlusion, occlusion, prime/compromised force sensor system and excessive no delivery test.No excessive no delivery/occlusion alarm noted.Device passed self test.Device received with cracked reservoir tube lip, cracked case at the reservoir tube window corners and cracked reservoir tube window.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|