| Model Number |
MMT-723LNAL |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.No button error alarm noted during testing.Unit had intermittent buttons response due to flattened (creased) down button dome switch.No unlocked j2/lcd keypad connector noted.Device passed rewind test and no rewind loud noise anomaly noted.Device had cracked battery tube threads, broken belt clip slot.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had keypad anomaly.The customer¿s blood glucose level was unknown.The customer reported that the buttons were less responsive and were harder to press sometimes had to press buttons twice.Customer stated that battery compartment had piece broken and flaking off hen changing battery every 2 weeks.Customer stated that and slower during rewind and louder during rewind.The insulin pump will not be returned for analysis.
|
| |
|