• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC LEFT THORACIC TACTILE PROBE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC LEFT THORACIC TACTILE PROBE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9734436
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Name of initial reporter unknown at the time of filing.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system.It was reported that the site found that the instrument tip was bent while cleaning the it.There was no patient present when this issue was identified.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received additional information that the instrument could not be verified.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received clarification that the damaged instrument was discovered by a manufacturer representative.
 
Additional Manufacturer Narrative
The probe was returned to the manufacturer for analysis.Analysis found that the tip of the returned probe was bent.However, with a tracker attached, the probe returned good geometry and divot error readings with normal tracking.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEFT THORACIC TACTILE PROBE
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9416631
Report Number1723170-2019-05875
Device Sequence Number11407957
Product Code OLO
UDI-Device Identifier00763000121228
UDI-Public00763000121228
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9734436
Device Catalogue Number9734436
Device Lot Number170907
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2019
Supplement Date Received by Manufacturer12/06/2019
01/02/2020
02/07/2020
Initial Report FDA Received Date12/05/2019
Supplement Report FDA Received Date12/27/2019
01/22/2020
02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-