| Model Number |
9734436 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Name of initial reporter unknown at the time of filing.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system.It was reported that the site found that the instrument tip was bent while cleaning the it.There was no patient present when this issue was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received additional information that the instrument could not be verified.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received clarification that the damaged instrument was discovered by a manufacturer representative.
|
| |
|
Additional Manufacturer Narrative
|
|
The probe was returned to the manufacturer for analysis.Analysis found that the tip of the returned probe was bent.However, with a tracker attached, the probe returned good geometry and divot error readings with normal tracking.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|