A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Device evaluation: visual analysis of the returned device identified: weight to the spec, crease flat, deformation.Cloudy and bubbles were observed after autoclave disinfection process.Based on the device analysis the final assessment is no issues found related with the manufacturing process.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iv.This is a known potential adverse event addressed in the product labeling.Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.
|