• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GNS II RESURF PAT 32MM; PRSTHSIS, KNEE, PTLLOFMRTIBIAL, SMI-CNSTRIND, CMNTED, POLYMER/MTL/PLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GNS II RESURF PAT 32MM; PRSTHSIS, KNEE, PTLLOFMRTIBIAL, SMI-CNSTRIND, CMNTED, POLYMER/MTL/PLYMR Back to Search Results
Model Number 71420576
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/27/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to pain and a feeling of unwillingness.
 
Additional Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation and the reported event could not be confirmed.The medical investigation concluded that, without the requested clinical information, a thorough medical investigation cannot be rendered.Should any additional medical information be provided, this complaint will be re-assessed.A review of complaint history did not reveal additional complaints for the listed batches.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Pain is a potential complication associated with any surgery.Some potential probable causes of this event could include patient reaction or a post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the devices or additional information be received, the complaints will be reopened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II RESURF PAT 32MM
Common Device Name
PRSTHSIS, KNEE, PTLLOFMRTIBIAL, SMI-CNSTRIND, CMNTED, POLYMER/MTL/PLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9419223
Report Number1020279-2019-04293
Device Sequence Number448573
Product Code JWH
UDI-Device Identifier03596010208309
UDI-Public03596010208309
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/02/2025
Device Model Number71420576
Device Catalogue Number71420576
Device Lot Number15KM02870
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/27/2019
Supplement Date Received by Manufacturer11/27/2019
Initial Report FDA Received Date12/05/2019
Supplement Report FDA Received Date04/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-