Visual inspection: x-ray images and the screw were examined and confirmed to have been fractured.Tip and shaft were reported to have fractured off and been left in the patient.Remaining portion of the screw was returned and showed little to no plastic deformation which indicates that brittle fracture had occurred.Device and complaint history records were reviewed, and no relevant manufacturing issues or similar complaints were identified.Per surgical technique: both the taps have a positive step at 10 mm.If preferred, self-starting screws may be used, which do not require any drilling or tapping.Screw insertion- insert the tifix bone screws through the plate using the size 10 tapered driver and tighten the bone screw until it begins to engage the plate.The size 10 tapered driver has a tapered, self-holding tip (stab and grab) to aid in insertion of the pyrenees screws.Repeat this procedure for the remaining screws."optional step" bone reduction- in cases where additional bone-lagging is required, a bone reduction screw can be used to draw the vertebral body closer to the plate.A contralateral tifix bone screw should then be inserted to hold the vertebral body to the plate.Replace the bone reduction screw with a tifix bone screw.Final tightening- after all the tifix bone screws have been secured, utilize the torque limiting handle in conjunction with the torque limiting hexalobe shaft for final tightening of the bone screws.The torque limiting screwdriver will provide a locking torque of 20 in-lbs (2.26 nm) and ensure the tifix locking technology is fully engaged and not over tightened.The anti-torque instrument is available for both pyrenees constrained and translational.This instrument docks onto the plate and is used as an anti-torque device when final tightening.Tifix locking technology- upon advancement of the screw, the screw head engages on the locking lip of the plate, lagging the plate down to the bone.Once bone lagging is complete, the revolutionary tifix locking technology commences.Due to a difference in material hardness and design, each screw head begins to reshape the screw hole and forms an autogenic lock to the plate upon insertion.With tifix, no additional locking is required.Realignment of a screw can be repeated up to three times without compromising the locking feature.It is unclear what caused the screw to fracture and a root cause for the reported issue could not be determined.
|