• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TISSUE SCIENCE LABORATORIES PERMACOL; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TISSUE SCIENCE LABORATORIES PERMACOL; MESH, SURGICAL Back to Search Results
Model Number P101015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Unspecified Infection (1930); Pain (1994); Hernia (2240); Injury (2348); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.Concomitant device: p151520 (lot#: 08b2209); p152030 (lot#: 09b0608); unk protack (lot#: unknown).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after implant, the patient experienced pain, infection, adhesions and recurrence.Post-operative patient treatment included revision surgery.
 
Additional Manufacturer Narrative
Additional info: a4, b7, d8, e1 (facility name, street, city, region, postal code), h6 (updated all codes, added patient codes and imf code).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERMACOL
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
TISSUE SCIENCE LABORATORIES
victoria house, victoria road
aldershot, hampshire GU11 1EJ
GB  GU11 1EJ
Manufacturer (Section G)
TISSUE SCIENCE LABORATORIES
victoria house, victoria road
aldershot, hampshire GU11 1EJ
GB   GU11 1EJ
Manufacturer Contact
justin ellis
60 middletown ave.
8200 coral sea st ne
mounds view,mn, CT 55112
7635265677
MDR Report Key9422400
Report Number9617613-2019-00198
Device Sequence Number19482158
Product Code FTM
UDI-Device Identifier10884523000115
UDI-Public10884523000115
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K120605
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2012
Device Model NumberP101015
Device Catalogue NumberP101015
Device Lot Number09B0701
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2019
Supplement Date Received by Manufacturer11/29/2023
Initial Report FDA Received Date12/06/2019
Supplement Report FDA Received Date12/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexMale
Patient Weight86 KG
-
-