| Catalog Number |
046001104 |
| Medical Device Problem Code |
Misassembled (1398)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Issue occurred prior to patient involvement.The involved device is not available for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Not available.
|
| |
|
Event or Problem Description
|
|
Livanova usa in has received a report that, during the test before initiating the procedure, it was found that the arterial and the venous lines were swapped they rectified it by switching the 3/8" tubing on the arterial and venous cannulas.The issue occurred prior to any patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The complained device was not made available for investigation.The customer has provided photographic evidence in which it is visible that the arterial and venous lines were swapped when compared with the technical drawing of the circuit.A review of the dhr did not identify any deviations, non-conformities or material scrap/requests relevant to the reported issue.No other similar complaint has been received for the claimed circuit and/or lot.Livanova investigation confirmed the reported condition.The most probable root cause of the claimed issue was an isolated human error during the manual assembly of this specific unit of this item circuit.To prevent further occurrences,along with the already present in process control, the involved operators will be retrained.Livanova will keep monitoring the market.
|
| |
|
Search Alerts/Recalls
|